3. Warnings and Precautions
Green Smart User Manual 13
EN
G
LISH
3.2
Electricity-related Safety Precautions
To avoid the risk of electric shock, this equipment must only be
connected to supply mains with protective earth.
Check the status of the power source, PC, and cables prior to operating the
equipment.
Ensure that Main Power Switch is set to off when the equipment is not in use.
Always disconnect the power supply before cleaning the equipment.
Always keep electrical cords away from hot appliances or radiators.
DO NOT place the PC or peripheral equipment connected to the PC near the patient.
The equipment and PC should be connected to common protective earth.
Never overload the equipment’s circuit by sharing it with too many appliances.
Use the same power circuit for the PC and the equipment.
Combining this equipment with other devices
Do not connect this equipment to devices which are not designated as a part of the
system.
Do not connect this equipment to a Multiple Portable Socket-Outlet (MPSO) or
extension cord which is not provided with the equipment.
Electromagnetic compatibility
This X-ray equipment complies with IEC standard 60601-1-2.
Medical electrical equipment is subject to special Electromagnetic Compatibility
(EMC) preventive measures. It must be installed and operated as specified in EMC
information.
If high-voltage systems, radio link systems or MRI systems are located within 5 m of
the unit, please observe the specifications stated in the installation requirements.
Portable Radio Frequency (RF) communications equipment may interfere with
medical electrical equipment. Therefore, the use of mobile wireless phones in
medical offices or hospital environments must be prohibited.
For more details, refer to 15.3 Electromagnetic Compatibility (EMC) Information.
Please also observe the Electro-Static Discharge (ESD) protective measures
described.
Summary of Contents for GREEN SMART
Page 1: ...User Manual Version 1 23 English...
Page 2: ......
Page 3: ......
Page 4: ......
Page 52: ...5 Imaging Software Overview 44 Green Smart User Manual Left blank intentionally...
Page 90: ...7 Acquiring PANO Images 82 Green Smart User Manual Left blank intentionally...
Page 110: ...8 Acquiring CEPH Images Optional 102 Green Smart User Manual Left blank intentionally...
Page 126: ...9 Acquiring CBCT Images 118 Green Smart User Manual Left blank intentionally...
Page 134: ...10 Acquiring 3D MODEL Scan Images 126 Green Smart User Manual Left blank intentionally...
Page 142: ...12 Cleaning and Maintenance 134 Green Smart User Manual Left blank intentionally...
Page 144: ...13 Disposing of the Equipment 136 Green Smart User Manual Left blank intentionally...
Page 154: ...14 Technical Specifications 146 Green Smart User Manual Left blank intentionally...
Page 185: ...15 Appendices Green Smart User Manual 177 ENGLISH...
Page 200: ......