9.6.8.GI-TL3 Compatibility output
(
GI-D7 Only
)
The table below gives the relationship between pressure and output voltage.
Table. 9-12
Measurement value outputs
Displayed pressure
value (Pa)
Output voltage
(V)
Displayed pressure
value (Pa)
Output voltage
(V)
6.00x10
-1
9.56
2.00x10
-3
4.60
4.00x10
-1
9.20
1.00x10
-3
4.00
2.00x10
-1
8.60
6.00x10
-4
3.56
1.00x10
-1
8.00
4.00x10
-4
3.20
6.00x10
-2
7.56
2.00x10
-4
2.60
4.00x10
-2
7.20
1.00x10
-4
2.00
2.00x10
-2
6.60
6.00x10
-5
1.56
1.00x10
-2
6.00
4.00x10
-5
1.20
6.00x10
-3
5.56
2.00x10
-5
0.60
4.00x10
-3
5.20
1.00x10
-5
0.00
Displayed pressure
value (Torr)
Output voltage
(V)
Displayed pressure
value (Torr)
Output voltage
(V)
6.00x10
-
3
9.56
2.00x10
-
5
4.60
4.00x10
-
3
9.20
1.00x10
-
5
4.00
2.00x10
-
3
8.60
6.00x10
-
6
3.56
1.00x10
-
3
8.00
4.00x10
-
6
3.20
6.00x10
-
4
7.56
2.00x10
-
6
2.60
4.00x10
-
4
7.20
1.00x10
-
6
2.00
2.00x10
-
4
6.60
6.00x10
-
7
1.56
1.00x10
-
4
6.00
4.00x10
-
7
1.20
6.00x10
-
5
5.56
2.00x10
-
7
0.60
4.00x10
-
5
5.20
1.00x10
-
7
0.00
P: Pressure value
(
Pa
)
V: Measured output voltage
(
V
)
P=10
(V/2-5)
(Pa)