53
When chose
WIB
Table. 9-11
Measurement value outputs
Displayed pressure
value (Pa)
Output voltage
(V)
Displayed pressure
value (Pa)
Output voltage
(V)
2.00x10
-1
5.20
1.00x10
-4
2.10
1.00x10
-1
5.10
5.00x10
-5
1.50
5.00x10
-2
4.50
1.00x10
-5
1.10
1.00x10
-2
4.10
5.00x10
-6
0.50
5.00x10
-3
3.50
1.30x10
-6
0.13
1.00x10
-3
3.10
5.00x10
-4
2.50
圧力表示値
(Torr)
出力電圧
(V)
圧力表示値
(Torr)
出力電圧
(V)
2.00x10
-3
5.20
1.00x10
-6
2.10
1.00x10
-3
5.10
5.00x10
-7
1.50
5.00x10
-4
4.50
1.00x10
-7
1.10
1.00x10
-4
4.10
5.00x10
-8
0.50
5.00x10
-5
3.50
1.30x10
-8
0.13
1.00x10
-5
3.10
5.00x10
-6
2.50
The measurement output voltage can be converted to a pressure value using the following
equation.
P: Pressure value
(
Pa
)
V: Measured output voltage
(
V
)
E: V from which decimal point is rounded down
(
V
)
P=10(V-E)
×
10
-(6-E)
(Pa)
Fig. 9-8 GI-D6 Compatibility output (WIB)
0
1
2
3
4
5
6
1.0E-06
1.0E-05
1.0E-04
1.0E-03
1.0E-02
1.0E-01
1.0E+00
O
u
tp
u
t V
ol
tage
(V
)
Pressure (Pa)
Summary of Contents for GI-M2