8.SETTING
8.1.Setting operation
To set sensitivity factor, set points, range changeover, degassing, filament, recorder
output and baud rate, make settings according to the explanations in 8.1 and higher.
These settings can be entered from the front panel or through the communication RS-
232C. Here, items to be entered from the front panel are explained.
A set status and set values are retained even after the power is turned off and are
effective at the next startup.
Holding down the [
○
PRG] key for 2 seconds or more from the POW-
①
or MESU-
①
status sets up the setting mode. (to SENS-
①
)
When the setting mode is set up, the LED above the [
○
PRG] key in the figure below
lights.
If the setting is not changed, press the [
○
PRG] key, If the change of the set point is
saved, press the [
⇒
(ENTER)] key to transfer control to the next.
To exit the setting mode, hold down the [
○
PRG] key for 2 seconds or more to revert to
the measurement mode, and the LED above the [
○
PRG] key will go off.
Inputs from the front panel are not feasible when in the [REMOTE] status. Always
reset them before input, because the keys on the front panel are locked when in the
[REMOTE] status.
If the [
⇒
(ENTER)] key is pressed after entering a numeric value out the settable range,
[
Err
12
] error will be displayed for about 1 second. Enter a settable value again.
Turn on in the
setting mode
(
Holding the key
)
Setting mode sets up / to exit press
the key for 2 seconds or more.
(
Normal Operation
)
Move to the next setting without save
it.
Blinking during setting
Save and move next
Change the operations
Change the value