68
11.4.5. Reading the emission valid signal
Command
Parameter
Response
Explanation of function
'
EM
'
-
'
OK
' or '
NG
' Reads emission valid status
'
ES
'
-
'
05
' or '
50
' Reads emission status (GI-N8)
'
E0
'
-
'
OK
' or '
NG
' Selects emission current 0.5mA (GI-N8)
'
E1
'
-
'
OK
' or '
NG
' Selects emission current 0.5mA(GI-N8)
Send back the emission valid. Send back '
NG
' when the filament OFF.
GI-N8 can read about setting and conditons of emission current and at the time, it points
at
“
05
”
:0.5mA
、“
50
”:
5mA.
11.4.6. Setting the sensitivity
→
Refer to
P.
35
Command
Parameter
Response
Explanation of function
‘SA’
-
‘OK’ or ‘NG’ Sets sensitivity factor (N
2
)
‘SB’
-
‘OK’ or ‘NG’ Sets sensitivity factor (A
r
)
‘SE’
‘
X.XX
E
±
XX
’ ‘OK’ or ‘NG’ Sets sensitivity factor
GI-M2
GI-D7
GI-N8
‘SA’
N2
N2
N2
‘SB’
Ar
Ar
H2
‘SE’
‘X.XXE
±
XX’
‘X.XXE
±
XX’
‘X.XXE
±
XX’
It is memorized, setting of value
‘
SE
X.XXE
±
XX
’ by the main body.
It is reflected at the time of the front panel / IO operation.
11.4.7. Setting and reading the set point
→
Refer to
P.
37
Command
Parameter
Response
Explanation of function
‘R1’
-
‘
X.XXE-XX
’ Reads data of set point-1
‘R2’
-
‘
X.XXE-XX
’ Reads data of set point-2
‘S1’
‘
X.XXE-XX
’ ‘OK’ or ‘NG’ Sets set point-1
‘S2’
‘
X.XXE-XX
’ ‘OK’ or ‘NG’ Sets set point-2
‘SP’
-
‘1-
X
/2-
X
’
Read the status of setpoint-1 and -2
Reading of setting and setting value of the set point.
It is reflected at the time of the front panel / IO operation.
The state of the set point is ON
:
1, OFF
:
0
is set by X of reply data `1-X/2-X'.
11.4.8. Setting analog output
→
Refer to
P.
41
Command
Parameter
Response
Explanation of function
‘LI’
-
‘OK’ or ‘NG’ Selects analog “LIN”
‘LG’
-
‘OK’ or ‘NG’ Selects analog “Pseudo-LOG”
‘LH’
-
‘OK’ or ‘NG’ Selects analog “REC-HOLD”
‘L4’
-
‘OK’ or ‘NG’ Selects analog “LOG” (GI-M2)
‘L6’
-
‘OK’ or ‘NG’ Selects analog “GI-D6” compatibility output” (GI-
D7)
‘L3’
-
‘OK’ or ‘NG’ Selects analog “GI-TL3” compatibility output”
(GI-D7)
‘LS’
‘E-
XX
’
‘OK’ or ‘NG’ Selects range hold linear output (range)
Set an analog output method.
The set point of the `LS' command sets the range which is ,` E-XX '.
It is reflected at the time of the front panel / IO operation.