33
8.12.Selection of Analog Function
→
Refer to
P.
41
Setting of analog function. The set point as follows.
GI-M2
GI-D7
GI-N8
rEC:1
Eeach-digit linear output mode
rEC:2
Pseudo-logarithmic mode
rEC:3
Range hold liniea mode
rEC:4
logarithmic mode
GI-D6 compatibility output mode
Not used
rEC:
5
Not used
GI-TL3 compatibility output mode
Not used
REC-①
The
○
portion in the left figure blinks
after the segment indicator displays [
rEC
__
] for about 1 second. Each time the [
∆
]
key is pressed, the
○
portion of the
segment indicator in the left figure displays
[
rEC
_1
]
⇒
[
rEC
_2
]
⇒
[
rEC
_
3
]
⇒
[
rEC
_1
]
・・
cyclically.
To use recorder LIN(LOG), press the [
⇒
(ENTER)] key, with [
rEC
_1
]
selected, to transfer control to the BPS-
①
status.
To use analog REC-HOLD, press the [
⇒
(ENTER)] key, with [
rEC
_3
]
selected, to transfer control to the REC-
②
status.
8.13. Setting of Analog Function
→
Refer to
P.
41
Set the REC-HOLD numeric values. The set point as follows.
GI-M2
GI-D7
GI-N8
Setting
range
-0
~
-8
-1
~
-7
-1
~
-9
Settable numeric values are sensitivity factor N2. This setting range will change if a
sensitivity factor other than N2 is designated or if sensitivity setting is entered.
REC-②
Each time the [
∆
] key is pressed, the
numeric value of the blinking portion
changes. When a desired value is
displayed, press the [
⇒
(ENTER)] key to
transfer control to the BPS-
①
status.