31
8.9. Selection Of DEGAS Function
→
Refer to
P.
39
Setting of DEGAS function
dGS:1
⇒
MANUAL
MODE
dGS:2
⇒
AUTO
MODE
DGS-①
The
○
portion in the left figure blinks
after the segment indicator displays [
d
GS
__
] for about 1 second. Each
time the [
∆
] key is pressed, the
○
position on the segment indicator
displays [
dGS
_1
]
⇔
[
dGS
_2
]
cyclically.
To use DEGAS MANUAL, press the [
⇒
(ENTER)] key, with [
dGS
_1
]
selected, to transfer control to the FIL-
①
status.
To use DEGAS AUTO, press the [
⇒
(ENTER)] key, with [
dGS
_2
]
selected, to transfer control to the DGS-
②
status.
8.10. Setting of DEGAS Function
→
Refer to
P.
39
Set the numeric value of DEGAS auto OFF.
Settable numeric values are 01 to 99 minutes.
DGS-②
The
○
portion in the left figure blinks.
Each time the [
∆
] key is pressed, the
numeric value of the blinking portion
changes. When a desired numeric
value is displayed, press the [
⇒
(ENTER)] key to transfer control to the
FIL-
①
status.
Summary of Contents for GI-M2