Wellwash Versa Main Parts
Liquid system diagram
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
35
Dispense valves
Dispense pump
Liquid level sensors
Liquid containers
Waste container
Wash head manifolds
Liquid valve manifold
Aspirate pump
Liquid level sensor
R
B
A
Figure 3–27.
Liquid system diagram
The instrument is equipped with a USB port for an external memory
device (Figure 3–26). You can transfer wash protocols from one
instrument to another of the same model with the USB memory device.
Note
It is recommended to format the USB memory device if export or
import fails. For more information, refer to Chapter 12:
“
Troubleshooting Guide
”.
▲
You can check the functionality of the USB memory device by inserting
it into the USB port for memory device (Figure 3–25). The
“
WELLWASH
” folder will be created.
The instrument is equipped with a USB port for an external PC
connection. The connection is used for remote control when connected,
for example, to a robotic system and for service use. A separate manual,
Thermo Scientific™ Wellwash™ Versa Remote Control Command Sets
(Cat. No. D08894), is available on request.
The plate carrier of the Wellwash Versa instrument supports both
portrait
and
landscape
orientation of the 96-well plate and
landscape
orientation of the 384-well plate (Figure 3–24). A plate clamp is
incorporated into the plate carrier to keep the plate in place during
processing. A sensor in the plate clamp senses the presence of a
microplate.
Liquid system
diagram
USB memory
device port
USB PC port
Plate carrier