Menus, Tabs and Parameters
Main menu
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
65
Note
If a 384-well plate with a 1x16 / 384 wash head is selected, you
can select only the
Flat
well shape.
▲
Note
If a 96-well plate with a 2x8 / 96 (cell wash) wash head is selected,
you can select only the
Flat
,
U
, and
C
well shapes.
▲
Note
The well shape has an effect on the aspirate height and aspirate
mode.
▲
Figure 6–36.
Well shapes
●
Well offset
– Adjust the X offset of a plate. Adjustment is required
primarily with a 384-well plate. The offset value ranges from
-1.5 mm to 1.5 mm in increments of 0.1 mm.
Caution
With 384-well plates, check that the offset limit is not too big
to prevent dispensing outside the wells.
▲
Use the
Up
and
Down
arrow keys to adjust the value.
Press the
F
3
(Preview) key to move the plate carrier and align the wash
head over the first column of a plate. Then press the
Up
or
Down
arrow key to move the plate carrier in steps of 0.1 mm.
Warning
Well offset has an effect on the residual volume in the well.
With U- and V-shaped wells, setting the well offset off center will
increase the residual volume.
▲
Note
With Flat bottom wells, setting the well offset off center may
decrease the residual volume. This feature should only be used with the
Normal aspirate mode.
▲
Flat
Flat
bottom
C
Flat bottom
with curved
edges
U
Round
bottom
V
V bottom
Star A
Flat bottom
with curved
edges and 8 ribs
Star B
Showing the
orientation
of the 8 ribs