Menus, Tabs and Parameters
Maintenance menu
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
91
3.
Select the priming channel using the
Down
arrow key and/or press
the
OK
button.
4.
Select the soak time using the
Right
arrow key and number keys.
The default is
15 min
. Press the
OK
button.
Priming, preparing the soak and soaking take place.
5.
Press the
F
2
key to stop the soaking if you want a shorter soak time
than the set soak time and to continue the procedure.
Priming is then carried out.
6.
Press the
OK
button when the procedure has finished.
Repeat the wash head cleaning procedure with deionized distilled
water if cleaning agent has been used.