Menus, Tabs and Parameters
Main menu
64
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
When you press the
OK
button, the wash head/plate list box appears.
The default wash head setting is set in the
Settings
menu. When a one-
column wash head (1x8 or 1x12) is used, the blank wash head
should always be on the left of the wash head arm (viewed from the
front).
Select the wash head and plate and press the
OK
button. Refer to
“Wash head configuration” on page 84 and “Creating a protocol” on
page 57.
Note
Calibrate the wash head(s) if not already calibrated.
▲
You can select the well shape and well offset on the
Well
row in the
Main
menu.
If you change the well shape, check the step parameters afterwards.
When you press the
OK
button, the
Well type
window appears.
The following settings are available:
●
Well bottom
– Select the well shape.
Different well shapes of strips and plates available are shown in
Figure 6-36. The default well shape format is
Flat
.
Well