Menus, Tabs and Parameters
Main menu
80
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
You can stop a protocol momentarily and then start the protocol again.
The
Pause
step has the following parameters:
●
Buzzer
– Set the buzzer on (= Yes) or off (= No). There is a beep
when the Pause
step starts. The default is
No
.
●
Comment
– You can add text, for example, instructions. There is
space for 128 characters.
To end the Pause step, press the
OK
button to continue protocol
execution. A three-second alarm is sounded.
The
Layout
step is used to predefine the strip selection for a protocol.
This step determines the area on which the next steps are processed.
The layout is
landscape
when a 1x8, 2x8, or 1x16 wash head is
selected and
portrait
for a 1x12 or 2x12 wash head. The strips are
numbered 1–12 if an nx8 wash head is used and marked A–H for an
nx12 wash head. The strips are numbered 1–24 if a 1x16 wash head is
used.
If there is no Layout step in the protocol, the processed area is according
to the Main >
Layout
row selection.
Note
Only the strips that are selected to the Layout step are saved with
the protocol. Strips selected using the
Layout
row are not saved.
▲
Pause
Layout