Menus, Tabs and Parameters
Maintenance menu
90
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
The procedure cleans the wash head according to a set of actions. Carry
out this procedure if the wash heads are dirty or clogged.
The instrument fills the priming vessel with a selected liquid and soaks
the wash head for a selected period of time.
Note
Use a suitable cleaning agent and repeat the
Clean
procedure with
deionized distilled water.
▲
1.
Select the
Clean wash head
row and press the
OK
button.
2.
Change the wash buffer to cleaning agent or deionized distilled
water. Press the
OK
button.
Clean wash head