Menus, Tabs and Parameters
Maintenance menu
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
99
5.
Insert the first tare weighed plate onto the plate carrier and press
the
F
1
key to continue.
Dispensing is carried out.
to
6.
Weigh the dispensed plate and enter the mass of the dispensed
liquid in grams using the number keys.
7.
Insert the second
tare weighed plate onto the plate carrier and press
the
F
1
key to continue.
Dispensing is carried out.