Maintenance
Cleaning liquid bottles
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
117
4.
Wipe both the plate carrier and the area surrounding it with a mild
detergent.
5.
Clean the plate carrier immediately when spillages have occurred on
or around the plate carrier. Be especially careful with the area close
to the plate clamp!
Figure 11–43.
Cleaning the plate carrier and the surrounding area
6.
Move the plate carrier back to its normal position once it has been
cleaned.
Empty the waste bottle before cleaning it according to disposal
regulations (see “Disposal of materials” on page 122).
The bottles must be cleaned regularly depending on the applications,
using a mild detergent.
The liquid bottles, tubes, and tube connectors can be autoclaved some
five to ten times (Figure 11–44). The liquid level sensors or the liquid
filters cannot be autoclaved (Figure 11–44).
Caution
Clean or replace the liquid filter of the intake tube regularly.
For example, back flush the liquid filter with detergent or use an
ultrasonicator.
▲
Cleaning liquid
bottles