Menus, Tabs and Parameters
Maintenance menu
96
Thermo Scientific Wellwash Versa User Manual
Thermo Fisher Scientific
the accuracy % for reference use only. The instrument uses linear
regression to calculate the dispense valve open time versus volume.
Note
Ensure that the wash head type corresponds to the wash head
setting in the
Settings
menu.
▲
Note
Before running the calibration, make sure that the liquid system
operates correctly, for example, that the liquid filters and wash head tips
are clean.
▲
Select the
Calibration
row in the
Maintenance
menu and press the
OK
button.
Starting the calibration