Parameter
Description
Waste level
You can monitor the liquid level in the waste container
when you have entered the following information:
•
Volume of liquid in the waste container at the
beginning of a sequence
•
Upper limit for the liquid level in the waste container
The
Remain Time
property for the waste reports the
time left until the liquid level is expected to reach the
upper limit. The Chromeleon software calculates the
time from the current flow rate and calculated
current liquid level.
•
Warning limit when you want to be informed about
the liquid level in the waste container