5
•
Installation
Vanquish Pumps (VH-P10-A-02) Operating Manual
Page 73
b) With your other hand, insert the tubing between the lever and
the rotor and wind it around the rotor.
c) Release the lever.
d) Check that the peristaltic tubing is properly inserted in the
tubing holders. If it is not, push the tubing into the holders.
Make sure that the tubing is not pinched or squeezed in the
holders.
e) Check that the right tubing connector rests on the tubing holder.
If it does not, push the tubing connector onto the holder.
2. In the autosampler, the seal wash line to the pump and the seal
wash inlet port (right tubing connector) are interconnected.
Disconnect the seal wash line (clear silicone tubing) from the seal
wash inlet port (right tubing connector).
Figure 22: Seal wash connections in the autosampler
3. Route the seal wash line from the seal wash outlet of the metering
device through the tubing guides to the peristaltic tubing in the
pump.
4. In the pump, connect the seal wash line to the tubing connector on
the free end of the peristaltic tubing.
No.
Description
1
Seal wash inlet port
(right tubing connector)
1a
Seal wash line to reservoir
2
Seal wash outlet on the
metering device head
2a
Seal wash line to pump
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...