7
•
Maintenance and Service
Page 182
Vanquish Pumps (VH-P10-A-02) Operating Manual
7.14
Updating the Pump Firmware
When
Updating the pump firmware might be required, for example, when a
new firmware version is released that adds functionality or solves
problems of a previous version.
Items required
Firmware version/Chromeleon Service Release as appropriate
TIP
When a new firmware version is released, the new version will be
included in the next available Chromeleon Service Release. The new
firmware will
not
be transferred automatically to the device when you
install the Chromeleon Service Release.
Preparations
•
Read the release notes provided with the firmware and/or
Chromeleon Service release.
•
Verify the following:
♦
The pump is connected in Chromeleon.
♦
All operations on the instrument (Chromeleon 7) or timebase
(Chromeleon 6.8) that includes the pump have been stopped.
The instrument or timebase is idle.
♦
The pump flow is turned off and the system pressure is down to
zero.
Follow these steps
1. Start the Chromeleon 7 Instrument Configuration Manager or the
Chromeleon 6.8 Server Configuration program.
2. Perform a firmware update from the
General
tab page in the
configuration dialog box for the pump. For details, refer to the
Chromeleon Help
.
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...