7
•
Maintenance and Service
Page 134
Vanquish Pumps (VH-P10-A-02) Operating Manual
7.6
Y-Connector
When
Damage of Y-connector
Parts required
Y-connector
Tools and additional items required
•
Solvent suitable for purging to flush out harmful substances
•
Solvent suitable for your application
Preparations
1. To flush out harmful substances, purge the pump with an
appropriate solvent.
2. Turn off the pump flow and wait until the system pressure is down
to zero.
Follow these steps
Figure 40: Y-Connector, used as left Y-connector (A) or right Y-connector
(B)
No.
Description
1
Capillaries to outlet check valves on pump head
2
Capillary to purge unit (port marked
R
or
L
, depending on Y-connector
position)
1. Disconnect the Y-connector capillaries on the purge unit and on the
outlet check valves of the pump head.