Troubleshooting
Page 214
Vanquish Pumps (VH-P10-A-02) Operating Manual
8.5.4
Testing the Inline Filter/Static Mixer for Permeability
When
If you observe pressure pulsation, inconstant pressure, or high
backpressure at the column and pump
Additional items required
•
Water as solvent
•
Tissue
Follow these steps
•
The pump has an inline filter installed
Test the inline filter by following the steps in
•
The pump has the mixing system installed
Test the static mixer by following the steps in
section 7.10.1,
page 170