5
•
Installation
Page 74
Vanquish Pumps (VH-P10-A-02) Operating Manual
5. Locate the seal wash line for connection from the autosampler to
the seal wash reservoir.
6. Connect the seal wash line to the seal wash inlet port (right tubing
connector) on the autosampler.
7. Route the seal wash line from the autosampler to the solvent rack
through the tubing guides in the autosampler and all modules above
the autosampler in the system stack.
8. Route the seal wash line through the guide hole in the solvent rack.
Figure 23: Guide hole and tubing guides in the solvent rack
No.
Description
A
Guide hole
1+2
Not to be used for wash liquid line; reserved for other tubing
3
Tubing guide for wash liquid line
4
Not to be used for wash liquid line; reserved for other tubing
9. Prepare the seal wash line:
a) Feed the seal wash line through the retaining guide.
b) Feed the seal wash line through an open hole in the cap of the
seal wash reservoir. The retaining guide keeps the tubing in
place in the reservoir.
c) Close any open holes in the reservoir cap with cap plugs.
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...