5
•
Installation
Vanquish Pumps (VH-P10-A-02) Operating Manual
Page 69
Figure 19: Solvent line with shut off valve installed
10. Check the solvent lines across the entire flow path. Make sure that
the lines are not bent, pinched, or squeezed at any point in the flow
path.
5.6.6
Setting Up the Seal Wash System
The flow path of the seal wash system passes through the metering
device head in the autosampler and the pump heads in the pump. The
seal wash system comprises the following parts:
•
In the pump: Seal wash pump (peristaltic pump), seal wash lines,
and seal wash detector (drop detector)
•
In the autosampler: Seal wash lines
•
Reservoir for the seal wash liquid
Both, the autosampler and pump are shipped with the seal wash
components installed and the lines connected within each of the two
modules.
To complete the setup, interconnect the seal wash paths of the
modules, connect and fill the seal wash reservoir, and flush the seal
wash system. To complete the setup, follow the steps in
section 5.6.6.2,
page 70
5.6.6.1
Choosing the Seal Wash Liquid
Use 75% isopropanol in water and 0.1% formic acid (HPLC-grade;
preparation by volume, for example,
75 mL isopr 25 mL water +
0.1 mL formic acid
).
No.
Description
1
Shut off valve
2
Retaining guide
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...