7
•
Maintenance and Service
Page 132
Vanquish Pumps (VH-P10-A-02) Operating Manual
4. Check the seal wash lines over the entire flow path:
a) Make sure that no bending (kink), pinching or squeezing of the
seal wash lines is present at any point in the flow path.
b) If you have to cut tubing to length, use a tubing cutter. Make
sure that the cut is at right angle to the length of the line.
5. Flush the seal wash system (see
section 5.6.6.3, page 75
). Use fresh
seal wash liquid.
7.5.3
Replacing the Seal Wash Detector
When
•
Contamination of detector electrodes
•
Impaired functionality of the seal wash system
Parts required
Seal wash detector assembly, consisting of the detector and funnel
Follow these steps
Figure 38: Removing the seal wash detector assembly
1. Disconnect the tubing from the seal wash detector (port labeled
in
).
2. Unplug the detector cable from the
Drop Det
port.
3. Grasp and pull the seal wash detector assembly upward by the
funnel.
No.
Description
1
Funnel, seal wash detector
2
Seal wash detector
3
Detector cable
4
Seal wash line
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...