7
•
Maintenance and Service
Page 192
Vanquish Pumps (VH-P10-A-02) Operating Manual
NOTICE
Shipping the slide-in module improperly leads to damage on
the device. Always ship the slide-in module as described in this
operating manual.
CAUTION—Possible Contamination
Hazardous substances may have contaminated the device during
operation and may cause personal injury to service personnel.
•
Decontaminate all parts of the device that you want to return for
repair.
•
Fill in and sign the Decontamination Certificate, which is part of the
Service Return Form. Sign the certificate to confirm that the device
has been properly decontaminated and that it is free of hazardous
substances.
•
Thermo Fisher Scientific refuses to accept devices for repair if the
Decontamination Certificate is missing.
7.17.2
Installing the Slide-In Module
Tools required
Screwdriver, Torx T20
Preparations
1. Unpack the replacement pump. Remove the slide-in module from its
enclosure as described in the
Removing the Slide-In Module
section
in this operating manual.
2. Verify that the pump enclosure into which the slide-in module shall
be installed is clean. If required, clean the inner and outer surfaces
of the enclosure. See the
Cleaning the Pump section
in this operating
manual.
3. When installing the slide-in module to an enclosure in the system
stack, check that the enclosure is placed correctly in the stack.
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...