8
•
Troubleshooting
Page 198
Vanquish Pumps (VH-P10-A-02) Operating Manual
Firmware Failure
If a firmware failure occurred during operation of the module, an
exception log has been created about the processes during the firmware
failure. The firmware sends the exception log to the Audit Trail when the
module is connected in Chromeleon.
In this case, observe the following:
•
Send the Audit Trail as
.cmbx
file to the Technical Support before
you clear the log.
•
To clear the exception log and continue operation of the module,
perform the command
ExceptionLogClear
.
For more information, refer to the
Chromeleon Help
.
Chromeleon Tests
With Chromeleon 7.2.6 and later, tests are available to help you to
identify and eliminate the source for a problem (see
section 8.3,
page 207