7
•
Maintenance and Service
Vanquish Pumps (VH-P10-A-02) Operating Manual
Page 161
7.7.9
Testing the Piston Seals for Leakage
When
Pressure pulsation, shift in retention times and/or poor reproducibility
can indicate possible piston seal leakage.
If messages regarding piston seal leakage appear in the chromatography
software, for example, "The piston seal leakage has exceeded the
recommended limit (code 4121)".
Chromeleon 7.2.6 and later
Run the Advanced Leak Test (see
Chromeleon 6.8 and Chromeleon 7 earlier than Chromeleon 7.2.6
Parts and additional items required
•
Backpressure capillary (from Diagnostics tool kit) (optional)
•
Isopropanol
Follow these steps
1. From Chromeleon, start a seal wash cycle.
2. Disconnect the tubing from the seal wash detector (port labeled
in
).
Figure 66: Seal wash detector
3. Remove some liquid from the tubing, for example, by shaking the
tubing.
4. Set up the system in such a way that approximately 120 MPa of
backpressure is produced.
For example, install an appropriate backpressure capillary on the
pump outlet (outlet of the inline filter or static mixer).
No.
Description
1
Seal wash detector
2
Tubing connection port (port labeled
in
)
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...