Section 4
4-8 ThermoFlex
Thermo Scientific
diA9
SP
SEtuP
enter
mode
mode
mode
mode
xx.x C
xx.x C
xx
xx
Hi t
UnitS
Lo t
Lo t
xx
xx
xx
xx
Hi P1
Lo P1
Hi P1
dELAY
enter
enter
enter
enter
enter
enter
enter
enter
enter
Hi t
UnitS
degx
Lo P1
Figure 4-6
Setup Loop (All Chillers)
or
or
Continued on next page.
•
UnitS
are the temperature, fluid flow (only chillers with an
optional flow transducer) and pressure display scales.
Scales:
°C or °F
Defaults:
°C
GPM or LPM
GPM
PSI, Bar or kPa
PSI
•
Hi t
is the fluid's High Temperature alarm limit.
Range:
+3°C to +42°C
Default:
+42°C
Exceeding this limit flashes
Hi t
and, if enabled, sounds the
alarm. The chiller reaction depends on the alarm configuration
(see
ALr
on next page).
•
Lo t
is the fluid's Low Temperature alarm limit.
Range:
+3°C to +42°C
Default:
+3°C
Falling below this limit flashes
Lo t
and, if enabled, sounds
the alarm. The chiller reaction depends on the alarm
configuration (see
ALr
on next page).
•
Hi P1
is the pump's High Pressure discharge alarm limit.
T 1 T 0 Pump
Range:
3 to 100 PSI
Default:
100 PSI
T 5 Pump
Range:
2 to 105 PSI
Default:
105 PSI
P 1 P 2 Pump
Range:
3 to 100 PSI
Default:
100 PSI
Exceeding this limit flashes
Hi P1
and, if enabled, sounds
the alarm (see
Sound
on next page).
•
dELAY
is the length of time the pump can exceed the
Hi P1
alarm limit.
Note
This feature is only active if the
chiller is configured to shut down with a pressure alarm.
P
1
,
P
2, T
0
and
T
1 Range:
0 to 30 seconds
Default:
0 seconds
P
3 - P
5, T
5 Range:
0 to 60 seconds
Default:
0 seconds
Exceeding this limit flashes
Hi P1
and, if enabled, sounds
the alarm. The chiller reaction depends on the alarm
configuration (see
ALr
on next page).
P 3 Pump 60Hz
Range:
3 to 46 PSI
Default:
46 PSI
P 3 Pump 50Hz
Range:
3 to 32 PSI
Default:
32 PSI
P 4 Pump 60Hz
Range:
3 to 85 PSI
Default:
85 PSI
P 4 Pump 50Hz
Range:
3 to 60 PSI
Default:
60 PSI
P 5 Pump 60Hz
Range:
3 to 87 PSI
Default:
87 PSI
P 5 Pump 50Hz
Range:
3 to 56 PSI
Default:
56 PSI
T 9 Pump 50Hz
Range:
3 to
105 PSI
Default:
105
PSI
Summary of Contents for ThermoFlex10000
Page 2: ......
Page 36: ...ThermoFlex900 5000 Thermo Scientific...
Page 42: ...DI SDS 4 0 gpm 15 1 lpm ThermoFlex900 5000 Thermo Scientific...
Page 79: ...517 2014 16 2014 ThemoFlex24000...
Page 100: ...Section 2 2 20 ThermoFlex Thermo Scientific...
Page 136: ...Section 4 4 14 ThermoFlex Thermo Scientific...
Page 168: ...Section 7 7 14 ThermoFlex Thermo Scientific...
Page 191: ...C 10 ThermoFlex Appendix C Thermo Scientific...
Page 208: ......
Page 209: ......