Section 3
ThermoFlex 3-19
Thermo Scientific
Initial Filling
Ensure the reservoir drain plug on the back of the chiller is in place, or the
Riton fitting is closed, and that all plumbing connections are secure.
Before using any fluid refer to the manufacturer’s SDS for handling
precautions.
The reservoir has a sight tube and ball for easy fluid level monitoring.
Slowly
fill the reservoir with clean process fluid through the funnel only, failure to
comply may result in internal spillage.
Note
Filling the reservoir above
MAX LEVEL
fill line will result in an
over flow error (
O FLO
) causing the chiller to shut down.
Since the reservoir capacity may be small compared to your application and
air may need to be purged from the lines, have extra cooling fluid on hand to
keep the system topped off when external circulation is started.
Figure 3-7
Reservoir Cap
Before replacing the reservoir cap ensure the reservoir sight tube
ball stopper is securely in place, see next page.
Replace the reservoir cap by screwing it clockwise. Cap should be hand
tight.
Reservoir Sight Tube & Ball
MAX
LEVEL
MIN
LEVEL
Figure 3-8
Reservoir Sight Tube & Ball
WARNING
Locate and remove the reservoir cap by
unscrewing it counterclockwise.
To prevent the introduction of particulates
into the system, fill the chiller with the
reservoir bag filter in place.
CAUTION
Summary of Contents for ThermoFlex10000
Page 2: ......
Page 36: ...ThermoFlex900 5000 Thermo Scientific...
Page 42: ...DI SDS 4 0 gpm 15 1 lpm ThermoFlex900 5000 Thermo Scientific...
Page 79: ...517 2014 16 2014 ThemoFlex24000...
Page 100: ...Section 2 2 20 ThermoFlex Thermo Scientific...
Page 136: ...Section 4 4 14 ThermoFlex Thermo Scientific...
Page 168: ...Section 7 7 14 ThermoFlex Thermo Scientific...
Page 191: ...C 10 ThermoFlex Appendix C Thermo Scientific...
Page 208: ......
Page 209: ......