Section 3
ThermoFlex 3-3
Thermo Scientific
* Refer to Appendix A for country specific ratings.
Electrical Service Requirements (Standard chillers):
ThermoFlex900
Voltage ±10%
Frequency
Phase
Branch Circuit
Requirements
Line Cord
Plug
100 VAC
50 Hz
1Ø
15A
5-15P
115 VAC
60 Hz
1Ø
15A
5-15P
200 VAC
50 Hz
1Ø
15A
6-15P
208-230 VAC
60 Hz
1Ø
15A
6-15P
230 VAC
50 Hz
1Ø
*16A
1
, 15A
2
, 13A
3
-
ThermoFlex1400
Voltage ±10%
Frequency
Phase
Branch Circuit
Requirements
Line Cord
Plug
100 VAC
50 Hz
1Ø
20A
5-20P
115 VAC
60 Hz
1Ø
20A
5-20P
200 VAC
50 Hz
1Ø
15A
6-15P
208-230 VAC
60 Hz
1Ø
15A
6-15P
230 VAC
50 Hz
1Ø
*16A
1
, 15A
2
, 13A
3
-
ThermoFlex2500
Voltage ±10%
Frequency
Phase
Branch Circuit
Requirements
Line Cord
Plug
200 VAC P 1, P 2 Pump
50 Hz
1Ø
15A
6-15P
208-230 VAC P 1, P 2 Pump
60 Hz
1Ø
15A
6-15P
200 VAC T 1 Pump
50 Hz
1Ø
20A
6-20P
208-230 VAC T 1 Pump
60 Hz
1Ø
20A
6-20P
230 VAC
50 Hz
1Ø
*16A
1
, 15A
2
, 13A
3
-
ThermoFlex3500/5000
Voltage ±10%
Frequency
Phase
Branch Circuit
Requirements
Line Cord
Plug
200 VAC P 1, P 2 Pump
50 Hz
1Ø
15A
6-15P
200 VAC T 1, P 3 Pump
50 Hz
1Ø
20A
6-20P
200 VAC P 4 Pump
50 Hz
1Ø
30A
L6-30P
208-230 VAC P 1, P 2 Pump
60 Hz
1Ø
15A
6-15P
208-230 VAC T 1, P 3 Pump
60 Hz
1Ø
20A
6-20P
208-230 VAC P 4 Pump
60 Hz
1Ø
30A
L6-30P
230 VAC P 1 - P 4 Pump
50 Hz
1Ø
*16A
1
, 15A
2
, 13A
3
-
The chiller is intended for use on a dedicated outlet. The ThermoFlex has
an internal circuit protection that is equivalent (approximately) to the
branch circuit rating. This is to protect the ThermoFlex, and is not
intended as a substitute for branch circuit protection.
Continued on next page.
Summary of Contents for ThermoFlex10000
Page 2: ......
Page 36: ...ThermoFlex900 5000 Thermo Scientific...
Page 42: ...DI SDS 4 0 gpm 15 1 lpm ThermoFlex900 5000 Thermo Scientific...
Page 79: ...517 2014 16 2014 ThemoFlex24000...
Page 100: ...Section 2 2 20 ThermoFlex Thermo Scientific...
Page 136: ...Section 4 4 14 ThermoFlex Thermo Scientific...
Page 168: ...Section 7 7 14 ThermoFlex Thermo Scientific...
Page 191: ...C 10 ThermoFlex Appendix C Thermo Scientific...
Page 208: ......
Page 209: ......