Section 2
ThermoFlex 2-7
Thermo Scientific
• Pump curves are nominal values. Pressure values for centrifugal pumps are differential pressures
between the inlet and the outlet of the chiller.
• Pump performance results were obtained with no restrictions on the return to the system or with
any options installed. For example, utilizing the DI option will result in a 0.5 gpm flow reduction .
• Specifications obtained at sea level using water as the recirculating fluid, at a 20°C process
setpoint, 25°C ambient condition, at nominal operating voltage. Other fluids, fluid temperatures,
ambient temperatures, altitude or operating voltages will affect performance. See Section 3.
• Thermo Fisher Scientific reserves the right to change specifications without notice.
Pressure
100
80
60
40
20
7
6
5
4
3
2
1
20
40
60 80
100 LPM
5
10
15
20 25 GPM
Flow Rate
Bar PSID
A
B
C
D
A: P
4, 60
Hz
B: P
4, 50
Hz
C: P
3, 60
Hz
D: P
3, 50
Hz
Pumping Capacity
Centrifugal Pump P
3/P
4
Pumping Capacity
Positive Displacement Pump P
1/P
2
Magnetic Drive Pumps MD
1/MD
2
A: P
2/MD
2, 60
Hz
B: P
2/MD
2, 50
Hz
C: P
1/MD
1, 60
Hz
D: P
1/MD
1, 50
Hz
5
10 15
20 LPM
1
2
3
4
5 GPM
A
B
C
D
100
80
60
40
20
0
7
6
5
4
3
2
1
Bar PSIG
Pressure
Flow Rate
Summary of Contents for ThermoFlex10000
Page 2: ......
Page 36: ...ThermoFlex900 5000 Thermo Scientific...
Page 42: ...DI SDS 4 0 gpm 15 1 lpm ThermoFlex900 5000 Thermo Scientific...
Page 79: ...517 2014 16 2014 ThemoFlex24000...
Page 100: ...Section 2 2 20 ThermoFlex Thermo Scientific...
Page 136: ...Section 4 4 14 ThermoFlex Thermo Scientific...
Page 168: ...Section 7 7 14 ThermoFlex Thermo Scientific...
Page 191: ...C 10 ThermoFlex Appendix C Thermo Scientific...
Page 208: ......
Page 209: ......