Section 4
4-10 ThermoFlex
Thermo Scientific
Figure 4-6
Setup Loop (All Chillers)
•
StArt
is used to enable/disable the auto restart
function. When enabled the chiller will automatically
restart after a power failure or power interruption
condition.
Range:
on or oFF
Default:
oFF
NOTE
Consider any possible risks before enabling
this mode of operation.
•
CArE
is used to set the preventive care cleaning
frequency reminder for the chiller's air and fluid
filters, in hours. The time selected is based on your
operating environment, see Section 6.
Range:
off
Default:
L1
L1 (1000 hours)
L2 (2000 hours)
L3 (3000 hours)
Off disables the reminder. Exceeding this limit
flashes
FLtrS
, see Section 6.
Note
If your chiller is equipped with any of the
Optional Features
refer to the next page.
When the display indicates
StorE
press
to save
all
changes or press to not save
all
changes. The display will return either the process
fluid temperature or, if the chiller was off when you
entered the loop, a blank screen.
Continued from previous page
xx
on
oFF
Lx
CArE
AUto
StorE
StArt
CArE
enter
enter
enter
enter
Save all changes
Do not save all changes
Summary of Contents for ThermoFlex10000
Page 2: ......
Page 36: ...ThermoFlex900 5000 Thermo Scientific...
Page 42: ...DI SDS 4 0 gpm 15 1 lpm ThermoFlex900 5000 Thermo Scientific...
Page 79: ...517 2014 16 2014 ThemoFlex24000...
Page 100: ...Section 2 2 20 ThermoFlex Thermo Scientific...
Page 136: ...Section 4 4 14 ThermoFlex Thermo Scientific...
Page 168: ...Section 7 7 14 ThermoFlex Thermo Scientific...
Page 191: ...C 10 ThermoFlex Appendix C Thermo Scientific...
Page 208: ......
Page 209: ......