Section 4
ThermoFlex 4-11
Thermo Scientific
Figure 4-7
Setup Loop (Optional Features)
Continued on next page.
xx
xxx
xx
xx
FLt
indC
xx
di t
SEc
HZ
ALr
di t
FiLL
LoFLO
HiFLO
HiFLO
enter
enter
enter
enter
enter
enter
enter
enter
enter
OPT
HZ
CArE
•
HZ
is used to identify the incoming frequency for
chiller's with P 3 - P 5 pumps
and
the capability to run
on either 50 Hz or 60 Hz. The selected frequency
automatically adjusts the firmware's
fixed
high pressure
default setting.
Range:
50 Hz or 60 Hz
Default:
60 Hz
•
OPt
is used to configure the analog in/out mode of
operation. See Appendix C.
•
FiLL
is used to set the time limit the auto refill has for
filling the reservoir to the normal operating level.
Range:
0 to 900 seconds
Default:
45 seconds ThermoFlex900 - 5000
80 seconds ThermoFlex7500 - 24000
Exceeding the time limit flashes
rEFiL
and the auto refill
will shut off. The chiller's reaction depends on the alarm
ALr
setting,
Flt
is shut down,
indC
is continue to run.
Note
Setting the time limit to 0 disables the auto refill
option.
See Section 5 for additional information.
•
di t
is used to set the preventive care cleaning frequency
reminder for the chiller's DI filter cartridge.
Range:
0 to 9999 hours
Default:
448 hours
Exceeding the limit flashes
di
,
see Section 6.
•
HiFLO
is used to set the high flow alarm limit.
Exceeding a high limit flashes
HiFLO
and, if enabled,
sounds the alarm. The chiller's reaction depends on the
alarm (
ALr
) setting.
Note
This feature is not enabled until the value is
changed to something other than 0.0.
Shaded displays
only appear
on chillers equipped with that option.
T 0 T 1
Range:
0.0 to 10.5 GPM
Default:
0.0 GPM
T 5 Pump
Range:
0.0 to 15.0 GPM
Default:
0.0 GPM
P 1 Pump
Range:
0.0 to 10.5 GPM
Default:
0.0 GPM
P 2 Pump
Range:
0.0 to 10.5 GPM
Default:
0.0 GPM
P 3 Pump
Range:
0.0 to 30.0 GPM
Default:
0.0 GPM
P 4 Pump
Range:
0.0 to 30.0 GPM
Default:
0.0 GPM
P 5 Pump
Range:
0.0 to 30.0 GPM
Default:
0.0 GPM
T 9 Pump
Range:
0.0 to 33.5 GPM
Default:
33.5 GPM
Summary of Contents for ThermoFlex10000
Page 2: ......
Page 36: ...ThermoFlex900 5000 Thermo Scientific...
Page 42: ...DI SDS 4 0 gpm 15 1 lpm ThermoFlex900 5000 Thermo Scientific...
Page 79: ...517 2014 16 2014 ThemoFlex24000...
Page 100: ...Section 2 2 20 ThermoFlex Thermo Scientific...
Page 136: ...Section 4 4 14 ThermoFlex Thermo Scientific...
Page 168: ...Section 7 7 14 ThermoFlex Thermo Scientific...
Page 191: ...C 10 ThermoFlex Appendix C Thermo Scientific...
Page 208: ......
Page 209: ......