Acquiring APCI Sample Data Using the Tune Plus Window
Acquiring APCI Data in the SIM Scan Mode
Thermo Electron Corporation
LTQ XL Getting Started
137
Figure 68.
Divert/Inject Valve dialog box
b. Select the
Load
option button, and overfill the 5-
μ
L injector loop
with the 125 fg/
μ
L solution of reserpine (or a solution of your
analyte of interest).
c. Select the Inject option button.
10. Observe the reserpine peak (
m/z
609.2), or that of your analyte of
interest, in the Spectrum view.
11. Perform the following repetitive sequence to obtain a total of four
consecutive loop injections of reserpine in the SIM scan mode. Wait
about 1 min between injections.
a. Select the Load option button to put the divert/inject valve in the
Load position. Overfill the injector loop with the 125 fg/
μ
L
solution of reserpine.
b. Select the Inject option button to inject the reserpine solution into
the APCI source. Then, observe the Spectrum view.
c. Wait 1 min before the next injection.
d. Repeat
through
three times
12. Click
Stop
in the Acquire Data dialog box to end the data acquisition.
Review the mass spectrum and chromatogram in the raw file you just
acquired using the Xcalibur Qual Browser window. See
For more information about reviewing the data you acquire using the LTQ
XL MS detector with the Xcalibur data system, refer to the
Xcalibur Getting
Productive: Qualitative Analysis
manual.