Setting Up the Ion Source for Acquiring Data in APCI/MS/MS Mode
Installing the Ion Max Ion Source Housing
110
LTQ XL Getting Started
Thermo Electron Corporation
a. Connect the 1-in. ID Tygon tubing to the ion source housing drain
fitting.
b. Attach the free end of the hose to a waste container. Ideally, the
waste container should be vented to a fume exhaust system.
The Ion Max ion source housing is now properly installed on the MS
detector.