Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Installing the Ion Sweep Cone
34
LTQ XL Getting Started
Thermo Electron Corporation
Installing the Ion
Sweep Cone
The ion sweep cone is a metallic cone that is installed over the ion transfer
tube. The ion sweep cone channels the sweep gas towards the entrance of
the capillary. This helps to keep the entrance of the ion transfer tube free of
contaminants. The net result is a significant increase in the number of
samples that can be analyzed without a loss of signal intensity. In addition,
keeping the ion transfer tube entrance cleaner reduces the need for frequent
MS detector maintenance.
Install the ion sweep cone
1. Remove the ion sweep cone from its storage container. Inspect and clean
it if necessary.
2. Note the location of the sweep gas supply port in the API cone seal. The
gas inlet on the ion sweep cone is placed in this port. See
and
.
Figure 9.
Sweep gas supply port in the API cone seal
Sweep Gas
Supply Port