Enter Normal Mass Range Data into Tune Plus
150
LTQ XL Getting Started
Thermo Electron Corporation
Enter Normal Mass
Range Data into Tune
Plus
Enter Normal Mass Range Data into Tune Plus
1. After carrying out the normal mass range calibration open the Tune Plus
window (
Instrument Set Up>LXQ>Start Tune Plus
).
2. In the Tune Plus window, open the Diagnostics dialog box
(
Diagnostics>Diagnostics
) shown in
.
Figure 70.
Tune Plus Diagnostics dialog box, normal mass range data
3. Click
Tools
in the upper left of the window. Click on Mass Calibration
in the pane below the
Tools
button as shown in
.
4. Items mentioned in
through
below are inside of the
Manual Coarse Calibration group box.
5. In the Calibrate Current Scan Type group box, choose User in the Mass
list box. In the Expected Low Mass box type 195.10 and type 1822.00