7
Optimizing the MS Detector with Your Analyte in APCI/MS Mode
Setting Up the MS Detector for APCI/MS Operation
Thermo Electron Corporation
LTQ XL Getting Started
119
Setting Up the MS
Detector for APCI/MS
Operation
Set up the MS detector for APCI/MS operation
1. In Tune Plus, click
On/Standby
to take the MS detector out of Standby
mode and turn it On. The MS detector begins scanning, and the LTQ
XL MS detector applies high voltage to the corona needle and shows a
real-time display in the Spectrum view.
2. Open the
APCIhighflow.LTQTune
Tune Method, the Tune Method for
high-flow APCI operation, as follows:
a. Choose
File > Open
to display the Open dialog box.
b. Scroll down until you see the folder
C:\Xcalibur\methods
. Then,
select the file
APCIhighflow.LTQTune
.
c. Click
OK
to open the file. LTQ XL MS detector downloads the
Tune Method parameters to the MS detector.
3. Verify that the LTQ XL MS detector opened the Tune Method, as
follows:
a. On the Instrument Setup toolbar, click
API Source
to open the
APCI Source dialog box. See
Figure 58.
APCI Source dialog box, showing the proper settings for a typical
high flow experiment
On
Standby