4
Tuning with Your Analyte in LC/ESI/MS Mode
Setting Up to Tune the MS Detector with Your Analyte
74
LTQ XL Getting Started
Thermo Electron Corporation
Setting Up to Tune the
MS Detector with
Your Analyte
Set up the MS detector to tune automatically on your analyte in ESI/MS
mode
In this procedure, you can use the reserpine solution described in
, or you can use a solution of an analyte of interest
to you.
1. Open the Tune Plus window from the Start button on your
Windows XP Desktop, as follows:
a. Choose
Start > Programs > Xcalibur > Xcalibur
to display the
Xcalibur Home Page – Roadmap view.
b. Click the Instrument Setup button to display the Instrument Setup
window.
c. Click the LTQ XL button to display the New Method page.
d. Click the Tune Plus button to display the Tune Plus window.
2. In Tune Plus, click the
On/Standby
button to take the MS detector out
of Standby mode and turn it On. The MS detector begins scanning, the
LTQ XL MS detector applies high voltage to the ESI probe, and the
LTQ XL MS detector shows a real-time display in the Spectrum view.
3. Open the ESImyTune.LTQTune Tune Method, the Tune Method you
saved in Chapter 3, as follows:
a. On the File/Display toolbar, click the
Open File
icon to display the
Open dialog box.
b. Browse to the folder C:\Xcalibur\methods. Then, select the file
ESImyTune.LTQTune.
c. Click
Open
to open the file. Tune Plus downloads the Tune Method
parameters to the MS detector, and the title bar in the Tune Plus
window should read as follows:
C:\Xcalibur\methods\ESImyTune.LTQTune – Tune Plus
CAUTION
Do not infuse calibration solution at flow rates above 10
μ
L/min. Ultramark 1621 can contaminate your system at high
concentrations.
Note
The following procedures assume that you are familiar with your
LTQ XL instrument and the Tune Plus window. If you need additional
guidance, see LTQ XL online Help and/or the
LTQ XL Hardware
Manual.
On Standby