Setting Up the Ion Source for Acquiring Data in APCI/MS/MS Mode
Removing the ESI Probe
104
LTQ XL Getting Started
Thermo Electron Corporation
counter-clockwise to free the probe from the alignment notch. Be
careful not to break the fused-silica sample tube or PEEK safety sleeve.
9. Pull the probe straight out to remove it from the ion source housing.
10. Store the ESI probe in its original shipping container.
The ESI probe is properly removed from the Ion Max ion source housing.