Tuning and Calibrating Automatically in the ESI/MS Mode
Setting Up the MS Detector in the Xcalibur Data System for Tuning and Calibration
50
LTQ XL Getting Started
Thermo Electron Corporation
a. On the Control/Scan Mode toolbar, click the
Define Scan
button
to open the Define Scan dialog box. See Figure 20. (If your dialog
box appears different from the one shown in the figure, it is
probably because the advanced settings are not displayed. You can
turn on the advanced settings as follows: In Tune Plus, choose
ScanMode
, and then click
Advanced Scan Features
to select the
option.)
b. In the Scan Description group box, in the Mass Range list box,
select Normal to allow for a selection of mass ranges between m/z
150 to 2000.
c. In the Scan Rate list box, select Normal to specify a normal scan
rate.
d. In the Scan Type list box, select Full specify a full scan.
e. In the Scan Time group box, in the Microscans spin box, enter
1
to
set the total number of microscans to 1.
f. In the Max. Inject Time spin box, enter 200.000 to specify a 200 ms
maximum injection time.
g. In the Source Fragmentation group box, confirm that the On check
box is not selected (
) to specify that the ion source fragmentation
option is turned off.
h. In the Scan Ranges group box, in the Input list box, select From/To
to make available the First Mass and Last Mass text boxes in the
Scan Ranges table.