5
Acquiring ESI Sample Data Using the Tune Plus Window
Acquiring MS Data in the SIM Scan
Thermo Electron Corporation
LTQ XL Getting Started
97
a. Select the Load option button, and overfill the injector loop with
the 125 fg/
μ
L solution of reserpine.
b. Select the Inject option button to inject the reserpine solution into
the ESI source, and then observe the Spectrum view.
c. Wait 1 min before performing the next injection.
d. Perform
through
three more times.
14. Click
Close
in the Divert/Inject Valve dialog box to return to the Tune
Plus window.
15. Click
Stop
in the Acquire Data dialog box to end the data acquisition.
Click
Cancel
to close the Acquire Data dialog box.