Setting Up the Ion Source for Acquiring Data in APCI/MS/MS Mode
Removing the ESI Probe
Thermo Electron Corporation
LTQ XL Getting Started
103
Figure 50.
Ion Max ion source housing with ESI probe installed
4. Disconnect the Auxiliary Gas fitting (green) from the auxiliary gas inlet
(A) on the probe manifold. (
.)
5. Disconnect the Sheath Gas fitting (blue) from the sheath gas inlet (S) on
the probe manifold.
6. Remove the stainless steel ZDV fitting (Grounding Union) from the
grounding bar on the ion source housing.
7. Unlock the probe locking lever by rotating the lever open to its widest
position.
8. Carefully pull the probe straight back in the port in the housing until it
meets with the slot in the API interlock block. The guide pin on the
probe manifold will prevent you from rotating the probe until the pin is
aligned with the slot in the API interlock block. Once the probe is all
the way back and aligned with the slot, turn the probe 45 degrees
Sheath
Gas
Fitting
Sample
Transfer
Tubing
Grounding
Bar
Grounding
Union
8 kV
Cable
ESI
Interlock
Socket
Auxiliary
Gas
Fitting
Sample
Inlet
PEEK
Safety
Sleeve
Probe
Locking
Lever