Tuning and Calibrating Automatically in the ESI/MS Mode
Tuning the MS Detector Automatically in the ESI/MS Mode
56
LTQ XL Getting Started
Thermo Electron Corporation
Tuning the MS
Detector
Automatically in the
ESI/MS Mode
You tune the MS detector automatically in the ESI/MS mode to optimize
important parameters, including heated capillary voltage and tube lens
voltage.
Tune the MS detector automatically
1. On the Control/Scan Mode toolbar, click the
Tune
button to display
the Tune dialog box.
2. If necessary, click the
Automatic
tab to display the Automatic tuning
page. See
.
3. In the What to Optimize On group box, select the Mass option button
to make active the Mass spin box.
Figure 23.
Tune dialog box, showing the Automatic tuning page