7
Optimizing the MS Detector with Your Analyte in APCI/MS Mode
Setting Up the Inlet for Tuning Using High-Flow Infusion
116
LTQ XL Getting Started
Thermo Electron Corporation
Setting Up the Inlet
for Tuning Using
High-Flow Infusion
Plumbing connections for APCI/MS sample introduction from the syringe
pump into the LC solvent flow
1. Connect a 4 cm (1.5 in) segment of Teflon tubing with a (brown)
fingertight fitting and a (brown) ferrule to the (black) LC union. See
.
Figure 56.
APCI/MS plumbing connections for the syringe pump
2. Load a clean, 500-
μ
L Unimetrics syringe with 450
μ
L of a 125 fg/
μ
L
solution of reserpine or your analyte of interest. (See
for a procedure for making the reserpine tuning
solution.)
3. Insert the needle of a syringe into the segment of Teflon tubing and
place the syringe in the syringe holder of the syringe pump.
4. Connect a fused-silica infusion line from the (black) LC union to the
(black) LC Tee union, as follows. See
a. Connect the infusion line with a (brown) fingertight fitting and a
(brown) ferrule to the free end of the LC union.
b. Connect the other end of the infusion line with a (red) fingertight
fitting and a (brown) ferrule to the side arm of the LC Tee union.
Teflon Tube
(P/N 00301-22803)
Ferrule
(P/N 00101-18196)
LC Union
(P/N 00101-18202)
Fingertight Fitting
(P/N 00101-18081)