Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Removing the APCI Probe
Thermo Electron Corporation
LTQ XL Getting Started
31
Figure 6.
Ion Max ion source housing, detail of components
6. Remove the 8 kV cable from the corona needle high voltage receptacle
as follows:
a. Unlock the cable by rotating the locking ring counter-clockwise.
b. Unplug the 8 kV cable from the corona needle high voltage
receptacle.
7. Remove the corona needle as follows:
a. Unlock the ion source housing door by turning the locks 90 degrees
so that the knobs are horizontal.
b. Open the ion source housing door.
c. Using pliers, grasp the corona needle and pull it straight out of the
corona needle contact. See
ESI
Interlock
Socket
API
Interlock
Block
Probe
Locking
Lever
Door
Locks
Ion Source
Housing
Door
Probe
Collar
CAUTION
AVOID INJURY.
The corona discharge needle is very sharp
and can puncture your skin. Handle it with care.