Preface
Standards and Directives
6
Heracell Vios 160i / 250i CR
Thermo Scientific
Typically, this CO
2
incubator has been designed for being installed and operated in the following areas:
• Laboratories for cytobiological and biotechnological experiments of safety levels L1, L2, and L3
• Medical-microbiological laboratories in accordance with DIN EN 12128
• Research labs in clinics and hospitals
This device is intended for professional use only.
This device is only to be operated indoors.
The device may only be operated by a laboratory technician or an instructed person.
A laboratory technician is a person with appropriate technical training or with many years of
experience in the relevant field of work who, on the basis of his or her knowledge and experience, is
able to recognize and avoid dangers that may emanate from laboratory equipment.
An instructed person is a person who has been instructed by a laboratory technician to avoid these hazards.
Incorrect Use
Do not use cell or tissue cultures in the device that are not in accordance with the regulations of safety
levels L1, L2, and L3. Do not use tissues, substances or liquids that:
• are easily ignitable or explosive,
• release vapors that form combustible or explosive mixtures when exposed to air,
• release poisons.
The Cell Locker Chamber is not for use in medical devices.
Standards and Directives
The device complies with the following standards and guidelines:
• Low Voltage Directive 2014/35/EU
• IEC 61010-1:2010+ AMD 2016, Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 1: General requirements
• IEC 61010-2-010:2019, Safety requirements for electrical equipment for measurement, control
and laboratory use - Part 2-010: Particular requirements for laboratory equipment for the heating
of material
• EMC Directive 2014/30/EU
• IEC 61326-1:2012, Electrical equipment for measurement, control and laboratory use - EMC
requirements, Part 1: General requirements
• IEC 60529:2.1:COR 2018 (IP54)
• FCC 47, Part 15, § 15.107, EMC regulation FCC
• FCC 47. Part 15, § 15.109, EMC regulation FCC
For other countries, the applicable national regulations are binding.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...