6
Handling and Control
User Configuration
6-34
Heracell Vios 160i / 250i CR
Thermo Scientific
• Workspace temperature,
• CO
2
concentration,
• O
2
concentration,
• relative humidity.
After conversion of the device, the device control system must be adjusted to the gas-tight screen option.
Setting the Gas Tight Screen
1. Press the
Menu
key.
2. Select the menu command
Options
.
• The selection dialog shown in
is displayed.
3. Select the menu command Gas tight screen.
• The input dialog shown in
is displayed.
Figure 6-39.
Setting the Gas Tight Screen
1. To toggle between two options:
• Press the + key.
or
• Press the - key.
2. To accept and save the change:
• Press the
Enter
key.
• The system returns to the
Options
menu.
Note
Malfunction:
The conversion to gas-tight screen causes a change of the control parameters.
Failure to set the gas-tight screen function in accordance with the actually installed door version
may impair the incubation performance.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...