12
Data Communication
Error Memory Structure
12-10
Heracell Vios 160i / 250i CR
Thermo Scientific
Overview of the Possible Error Messages in Hex Coding
Overview of the Possible Error Messages in Bit Coding
General Device Status, Temperature and CO
2
Control Loop
Hex Code
Description / Type
0x00
Temperature control loop
0x01
CO
2
control loop
0x02
O
2
control loop
0x07
Water level
0x08
General device status
Bit
General device status
0x0002
Device Door open too long
0x0004
Display fails to communicate
0x0008
Mainboard parameter implausible (EPROM faulty)
0x0010
Data logger defective (unit can continue to function)
0x0020
Error for sterilization / Steri-Run
0x0040
Power-down during Steri-Run
0x0080
Error auto-start
0x0100
ADC-test failed
0x0400
Fan error
0x1000
IR sensor system replaced (info)
0x2000
auto-start active (info)
0x4000
Sterilization active (info)
0x8000
Device error has occurred (info)
Bit
Error status of temperature control circuit
0x0001
Sensor breakage
0x0002
Actual value high
0x0004
Actual value low
0x0008
Actual value not plausible
0x0010
Calibration values too high/low
Bit
CO
2 control loop error status
0x0001
Sensor breakage
0x0002
Actual value high
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...