12
Data Communication
Overview of Incubator Parameters (Addresses 2xxx)
Thermo Scientific
Heracell Vios 160i / 250i CR
12-7
Reading Internal Function Parameters
*1) See table with notes on sterilization and autostart run status.
*2) Two bytes each per level.
*3) Cylinder A active (0x01), cylinder B active (0x02), pressure in cylinder A OK (0x10), pressure in
cylinder B OK (0x20).
*4) For other Informations on the Error Log, see Chapter 13.5.
*5) Place the reading pointer on the first entry, read up to 7 entries max.
*6) Send the next 7 entries. Set reading pointer automatically to next new entry, read up to 7 entries max.
*7) Re-send entries of last telegram. Can be used after communication error has occurred.
Address
Description
Remarks
2100
Status of run
*1)
and
Remaining run time
[hours:minutes]
Sterilization, date
and time of last start
25 bytes / decimal value
with the format
xx;+xxx:xx;xx.xx.xx;xx:xx
2105
Status of run
*1)
, current
CO
2
waiting period
[minutes:seconds]
auto-start, Date and time of last start
25 bytes / decimal value
with the format
xx;xx.x;+xxx:xx;xx.xx.xx;xx:xx
2140
Status of as cylinder changeover
switch
Read CO
2
*3)
2 bytes / hexadecimal
with the format xx
2141
Status of as cylinder changeover switch
Read O
2
*3)
2 bytes / hexadecimal
with the format xx
2300
Read the Error Log
(current errors)
*4)
Up to 241 bytes / hexadecimal value
Format, see section on this topic
2301
Read the Error Log
(older errors)
*4)
Up to 241 bytes / hexadecimal value
Format, see section on this topic
2400
Query (Start) of data stored in the
data logger
*5)
Up to 224 bytes / hexadecimal value
Format, see section on this topic
2401
Query of other data stored in the
data logger
*6)
224 bytes / hexadecimal value
Format, see section on this topic
2402
(Repeat) query for last data logger
query
*7)
224 bytes / hexadecimal value
Format, see section on this topic
2410
Read out writing cycle of the data
logger in hours/minutes/seconds
8 bytes / decimal value with the
format xx:xx:xx
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...